13. Faults
If the measures listed below fail to remedy any faults or if the device has faults not
listed here, please contact Schaller GmbH.
Fault
Measuring error
Page | 53
Cause
The temperature is outside the
operating temperature: lower
than -10 °C or higher than
+60 °C
Measurement error due to too
short temperature adjustment
time
Wrong product type
Filling volume too low
Dripping water or sprayed
water
Irreversible damage of the sen-
sor due to aggressive gases
Condensation caused by a
change in temperature
Polluted sensor
Foreign particles on the sensor Please contact your dealer.
Remedy
Only use the device in tem-
peratures between -10 °C
and +60 °C.
Let the device adjust to
the surroundings (see
Adjustment behaviour of
the
sensor").
Check whether you have
selected the right product
type before taking a reading
(see
"7. Product
types").
Fill the screw-top jar to at
least two thirds with fine
material or at least half-way
with coarse material.
Direct contact of the sensor
with dripping or sprayed
water will destroy it.
Please contact your dealer.
Condensation on the sensor
interferes with the calibra-
tion. Let the device adjust
to the surrounding tempe-
rature.
Please contact your dealer.
"5.3