Schaller Messtechnik Humimeter AW3 User Manual page 52

Table of Contents

Advertisement

5.
Overwrite the current Ic-value with the new
calculated Ic-value for this index (figure 71). To do
so, press and hold
required number and either press it for 3 seconds
or press
(figure 72).
Moving back:
Press
to switch to another input level.
To move back, press
6.
Navigate to the next index by pressing
7.
Repeat steps 5 and 6 for the rest of the calculated
indices, leaving index7 (Idx:[ 7]) untouched.
8.
Press
to switch to another input level and
»
The recalibration has been saved.
9.
Press
to leave the Calibrate menu.
10. Press
to leave the Options menu.
11. Press
to leave the main menu.
12. In case of an incorrect recalibration, it is possible to reset the device to its factory
settings (see
»
By restoring the factory calibration, recalibrations that have already been carried
out successfully are deleted.
Page | 52
to quickly scroll to the
to confirm the selected number
.
"10.13 Resetting the device to its factory
71
72
twice.
to leave the Humidity menu.
settings").