The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it
Immunity Test
3 Vrms
150 kHz to 80 MH
3 V RMS outside the ISM band,
6 V RMS in the ISM and amateur bands
Conducted RF
80% AM at 1kHz
IEC 61000-4-6
10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
34
Guidance and manufacture's declaration – electromagnetic immunity
is used in such an environment.
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Not Applicable
Recommended separation distance
d=0.35√p
d=1.2√p