Fluids:
Only use the fluids, which fulfill the
CAUTION
requirements for safety, health and
device compatibility. Be aware of
the chemical hazards that may be associated with the
bath fluid used. Observe all safety warning for the fluids.
• Depending on the bath fluid used and the type of opera-
tion or toxic can arise. Ensure suitable extraction.
• Do not use any fluid which may cause dangerous reac-
tions during processing.
• Only use recommended bath fluid. Only use non-acid
and non corroding fluid.
NOTICE
at a regular basis (see the section "Filling and draining").
• Continuous monitoring of the bath and the filling level of
the bath fluid is required.
• To ensure a sufficient fluid circulation, the viscosity of
the bath fluid must not exceed of 50 mm
est operating temperature.
• Untreated tap water is not recommended. It is recom-
mended to use distilled water or high purity water (ion
exchangers) and add 0,1 g soda (sodium carbonate
Na
CO
) /liter, to reduce corrosive properties.
2
3
NOTICE
- Untreated tap water
- Acids or bases
- Solutions with halides: chlorides, fluorides, bromides,
iodides or sulfur
- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.
Refrigerant:
WARNING
WARNING
WARNING
• Set the device in well-ventilated room and keep all venti-
lation openings in the enclosure or, in the structure for
building, clear of obstruction.
Never operate the device without
sufficient refrigeration carrier flu-
id! Check the fluid level detection
2
/s at the low-
Don't use following fluids:
The device is not allowed to be
used in an ATEX (Atmosphere EX-
losible) zone.
Flammable refrigerant used. Do
not use mechanical devices or oth-
er means to accelerate the defrost-
ing process, other than those rec-
ommended by the manufacturer.
Do not damage the refrigerant cir-
cuit.
• Flammable refrigerant used. Only the manufacturer can
open the device and repair it. The third parties are not
allowed to open the device!
• The device with a sealed refrigerating system with less
than 150 g refrigerant. It can be located in any laboratory.
Battery pack RB 1 (for WiCo):
NOTICE
vice (station) will continue to run or is shut down depend-
ing on the value settings for
and
"Safe
Temperature". If the device is set so that it con-
tinues to run when the battery of the WiCo is fully dis-
charged, the only means of switching the station off are
the
"safe
STOP", "ON/OFF" key and the
NOTICE
• Keep the battery pack out of reach of children at all times.
• Store the battery pack in a cool, dry place.
• Never throw the battery pack into a fire. Keep it away from
direct sunlight and temperatures above 60 °C. High tem-
peratures will damage the battery pack and render it unus-
able. Temperatures above 100 °C may cause it to explode.
• Never throw the battery pack into water or expose it to
moisture. Water may lead to a short-circuit, causing the
battery pack to explode.
• Do not deform or crush the battery pack or damage it
in any other way. This can cause battery fluid to leak
and/or the battery pack to explode.
• When not in use, keep battery packs away from pa-
perclips, coins, keys, nails, screws or other small metal
objects which could cause the contacts to be bridged.
Short-circuiting may result in an explosion.
• Explosion of a battery pack may release battery fluid and
cause a fire.
• The lithium polymer battery pack must only be used and
charged in IKA products designed for use with this bat-
tery pack.
• When the battery pack is inserted it should slide in easily
and without resistance. Do not force it.
• If the battery pack is removed for an extended period of
time, store it in a sealed plastic bag to prevent short-circu-
iting due to moisture or contact with metal components.
• The operating temperature range of the battery pack is
from 0 °C to + 45 °C. Note that the battery pack capac-
ity will be reduced at temperatures below 20 °C.
• Only the rechargeable battery types recommended in the
technical data may be used in the device!
Do not charge batteries that have leaked or that
are discolored, deformed or damaged in any
other way.
If during operation the battery
pack RB 1 (rechargeable battery)
becomes fully discharged, the de-
"Time
Out",
"Safe
"Power
Please note the following
safety instructions for the
battery pack RB 1 (recharge-
able battery):
Speed"
switch"!
6