Nominal voltage
Frequency
Max. input power
Working temperature range
Operating temperature range (with external heating)
Temperature stability – internal temperature control
(according to DIN 12876)
Temperature control
Temperature measurement, absolute accuracy
Internal (int) (adjustable by calibration)
External (ext) (adjustable by calibration)
External Pt 100.3 temperature sensor tolerance to DIN EN
60751 class A, ≤ ± (0.15 + 0.002 x |T|), e.g. at max. 100°C
(adjustable by calibration)
Temperature setting
Temperature setting resolution
Temperature display
Temperature display resolution
Cooling capacity according to DIN 12876 (at 4600 rpm):
Refrigerant
Refrigerant quantity
Max. refrigerating system pressure
Pump speed (adjustable)
Max. pump pressure/suction
Max. flow rate (at 0 bar)
Bath volume
Fluid maximum viscosity
Low level protection
Interface
Permitted on-time
IP code according to EN 60 529
Protection class
Excess voltage category
Contamination level
Permitted ambient temperature
Permitted ambient humidity
Dimension (W x D x H)
Weight
Operation at a terrestrial altitude
Technical data
VAC
Hz
W
°C
°C
K
K
K
K
K
+ 20 °C
W
+ 10 °C
0 °C
- 10 °C
- 20 °C
- 30 °C
g
bar
rpm
bar
l/min
I
mm
2
/s
%
°C
%
mm
kg
m
230 ± 10% / 100–115 ± 10%
50 / 60
1100
- 30 ... RT
- 30 ... + 80
± 0.1
PID (Automatic / user setting)
± 0.5
± 0.5
± 0.35 (at 100°C)
Knob on WiCo
0.1
TFT LCD on WiCo
0.01
1400
1200
950
650
450
200
R 290
90
21
1000...4600
0.61 / 0.45
31
5–7
50
Yes
USB, RS 232, multi-function interface
100
IP 21
I
II
2
+ 5 ... + 32
80
310 x 490 x 546
37.8
max. 2000
29
Need help?