Storing Readings

1.
Press the MOLD button to store a reading. The
storage location number will be displayed on the
lower display, while the main display shows the
stored reading. The meter will enter the HOLD mode
and the "HOLD" indicator will appear.
2.
Press the MODE button again to exit the HOLD mode
and return to normal operation.
3.
If more than 25 readings are stored, previously
stored reading (starting with number 1) will be
overwritten.

Recalling Stored Readings

1.
Press the CAL button and then press the MODE button. The location
number (1 through 25) will briefly appear and then the value stored
in that location will appear. The displayed units will flash, indicating
that the storage recall mode is active.
2.
The last stored reading taken will be displayed first. To advance to
the previously stored readings, press the MODE button. The location
number is displayed first, followed by the reading stored in that
location.
3.
To exit the storage mode, press the CAL button and the TRACER will
return to normal operation after displaying "End".

Clearing Stored Memory

Turn the TRACER on. Press and hold the ON/OFF button for 4 seconds. The
display will briefly display "clr" when the memory is cleared.