New Spin on Precision Water Analysis Code 3576 Go to www.lamotte.com/spin for: Instructional videos • Testing Tips • Software Updates • Registering your meter...
Page 2
Borate, Biguanide, and Biguanide Shock. This innovative photometer is easy-to-use with a disk reagent system that holds the precise amount of reagents needed to run the tests. LaMotte chemists developed reagent disks specifically for the in-store pool and spa analyst who desired similar precision to current water labs but without the time-consuming procedures or testing limitations of test strips.
Failure to do so could result in personal injury or damage to the meter. The WaterLink Spin should not be stored or used in a damp or corrosive environment. Care should be taken to prevent water or reagents from entering the photometer chamber.
WARRANTY LaMotte Company warrants this instrument to be free of defects in parts and workmanship for 2 years from the date of shipment. If it should become necessary to return the instrument for service during or beyond the warranty period, contact our Technical Service Department at 1-800-344-3100, ext.
DISKS The WaterLink Spin uses a disk reagent system. The dried reagents are packaged in single test amounts in a sealed, polystyrene disk. Stainless steel mixing beads in the reaction chambers mix the sample water and the dried reagents.
Page 9
CHLORINE/BROMINE DISK Calcium Total Hardness, Chlorine High Range (Code 4328-H) Calcium Total Hardness Alkalinity Free Chlorine Calcium Hardness Bromine Copper Copper Iron Total Chlorine Iron Total Alkalinity Cyanuric Acid Free Chlorine/ Bromine Cyanuric Acid Blank CHLORINE/BROMINE PLUS PHOSPHATE Calcium Total Hardness, Chlorine High Range...
Wet disks should be dried thoroughly with a lint Under free wipe. Disks should be filled and used within Filled 10 minutes. They cannot be filled ahead of time. For filling and troubleshooting tips go to: www.lamotte.com/spin Under Filled...
The Calibration Check Disk (Code 1703) is used to verify the performance of the meter. See the Calibration Check Disk instructions for more information. USB CABLE The USB cable connects the WaterLink Spin to a computer. A USB 2 port is required. AC ADAPTER...
MAINTENANCE CLEANING The optical system of the Waterlink Spin must be kept clean and dry for optimal performance. Dry the disk with a lint-free wipe before placing it into the chamber to avoid introducing moisture. For best results store the instrument in...
Should it be necessary to return the meter for repair or servicing, pack the meter carefully in a suitable container with adequate packing material. A return authorization number must be obtained from LaMotte Company by calling 800- 344-3100, ext. 2 (US only) or 410-778-3100, ext. 2, faxing 410-778-6394, or emailing tech@lamotte.com.
TROUBLESHOOTING TROUBLESHOOTING GUIDE Problem Reason Solution Connection Error Could not connect to Adjust program water analysis program. settings. Be sure program is running. USB 2 port required. Lid Open Error Lid open. Close lid and begin again. Test Series Cancelled Lid open, meter not Close lid.
For precise Hardness results in sample water containing salt choose the Chlorine Generator product from the Sanitizer drop down bar in the water analysis program. Go to www.lamotte.com/spin for: • Instructional videos • Testing Tips • Software Updates •...
Page 16