99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com Tracer Fluoride PockeTester Code 1756 Water Testing Leader Since 1919!
Page 2
WARNING! This set contains chemicals that may be harmful if misued. Read cautions on individual containers carefully. Not to be used by children except under adult supervision.
Four (4) CR2032 button batteries Low Battery Indication ‘BAT’ appears on the LCD Auto Power Off After 12 minutes of inactivity Dimensions/Weight 1.4 x 6.8 x 1.6” (36 x 173 x 41 mm); 7.4 oz (210 g) LaMotte TRACER Fluoride PockeTester 1756 Instruction...
TISAB reagents and standards already in use. TISAB TesTabs contain all of the essential and approved chemicals that are found in the usual TISAB reagents. The “dry” reagent does not contribute to sample dilution. LaMotte TRACER Fluoride PockeTester 1756 Instruction...
Release the CAL key. Wait until the display stops blinking; the instrument will enter the HOLD mode. The instrument is now calibrated and ready for use or calibration with a second standard. LaMotte TRACER Fluoride PockeTester 1756 Instruction...
Wipe the f uoride crystal and reference junction with a damp tissue. (Cleaning techniques are addressed later in this guide). LaMotte TRACER Fluoride PockeTester 1756 Instruction...
For a 2 point calibration, repeat the calibration procedure with a 10.0 ppm standard. The instrument is now calibrated and ready for use. The circled ‘L’ and ‘H’ icons on the display indicate that the low range (L) and high range (H) calibrations have been completed. LaMotte TRACER Fluoride PockeTester 1756 Instruction...
ON. The CAL/RECALL key can be released when ‘Self Cal’ is shown in the display. To switch back to the previous unit of measure, repeat steps 1 through 3. LaMotte TRACER Fluoride PockeTester 1756 Instruction...
Clearing Stored Readings With the unit ON press and hold the ON/OFF key for 4 seconds. When ‘clr’ is shown in the main display the memory is cleared. LaMotte TRACER Fluoride PockeTester 1756 Instruction...
The instrument will give an error code when the electrode can no longer be calibrated. If the instrument will not calibrate, clean the f uoride electrode surface and recalibrate the instrument. If the meter still does not calibrate, replace the electrode. LaMotte TRACER Fluoride PockeTester 1756 Instruction...
WARRANTY LaMotte Company warrants this instrument to be free of defects in parts and workmanship for 1 year from the date of shipment and warrants this probe to be free of defects in parts and workmanship for 6 months from the date of shipment.