Output Scale; Output Signal Limitation - E+E Elektronik EE371 Manual

Dew point transmitter/switch down to - 60 c td
Hide thumbs Also See for EE371:
Table of Contents

Advertisement

2.4

Output scale; Output signal limitation

The Td analogue output can be scaled within the range -100...80 °C (-148...176 °F).
For air (gas) temperature Ta ≤20 °C (≤68 °F), the output is limited at Td = -80 °C (-112 °F)
For air (gas) temperature Ta >20 °C (≤68 °F), the output is limited at Td = Ta -100 °C (Ta-180 °F)
Example (see also Fig.1):
For air (gas) temperature Ta = 30 °C (86 °F) the output is limited at Td = 30°C -100 °C = -70°C (86°F –
180°F = -94 °F).
60
40
20
0
-20
-40
-60
Accuracy not specified
-80
-40
-30
Fig. 1
Dew point measurement range and specified accuracy
The maximum scaling of the analogue output is -100...80 °C (-148...176 °F) dew point.
Manual EE371 Dew point transmitter
Accuracy ≤ ±2 °C Td
-10
-20
0
10
Air (gas) temperature Td [°C]
20
30
50
40
Output signal
limitation
60
7

Advertisement

Table of Contents
loading

Table of Contents