Specifications
HM 5012-2 / HM 5014-2
Frequency
Frequency Range:
Stability:
Ageing:
Display Resolution (Readout):
Center Frequency Range:
Frequency Generation:
Span:
Zero span and 1MHz to 1000MHz in steps of 1-2-5
Marker: Resolution (Frequency)
Resolution (Amplitude)
Resolution Bandwidth, RBW (6dB):
Video Bandwidth, VBW:
Sweep time (autom. Selection):
Amplitude (Marker related) 150kHz - 1GHz
Measurement Range:
Scaling:
Display Range:
Frequency response (@10dB Attn., Zero Span, RBW 1MHz,
-20dBm Signal):
Display CRT, Units:
Input Attenuator Range:
Attenuator Accuracy rel. to 10dB:
Maximum Safe Input Level (continuous):
Attenuator setting 40dB:
Attenuator setting 0dB:
DC max:
Reference Level Variation:
Reference Level Accuracy (rel. to 500MHz, 10dB Attn., Zero Span,
RBW 1MHz):
Min. Average Noise Level:
Intermodulation (3rd Order):
(2 Signals, –27 dBm each, Frequency distance>3MHz)
Harmonic Distortion
(2nd at input level -27dBm, ATTN 0dB, Frequency
Resolution Bandwidth Amplitude Error (rel. to RBW 1MHz,
Zero Span):
Digitising Error:
±1 Digit (0.4dB) @ 10dB/div. (Average mode)
Inputs / Outputs
Signal Input:
Input Impedance:
Tracking Generator Out (HM5014)
Output Impedance:
Test Signal Output
Output Impedance, Frequency, Level:
Probe Power:
VSWR:
Audio Output (Phone):
Functions
Num. Keypad:
Reference and Tracking Generator Level
Rotary Dial:
Max. Hold Detection:
Quasi-Peak Detection:
Average:
SAVE/RECALL
AM-Demodulator
LOCAL:
Tracking Generator (HM5014-2 only)
Output Frequency Range:
Output Level:
Frequency Response:
Spurious Outputs:
28
(Reference Temperature 23°C ±2°C)
0.15 MHz to 1050MHz
±5ppm
±1ppm/year
1kHz (6½ -Digit)
0 MHz - 1050 MHz
TCXO with DDS
1KHz, 6½ -Digit
0.4dB, 3½ -Digit
9kHz, 120kHz and 1MHz
40ms, 320ms, 1s*
–100dBm to +10dBm
10dB/div., 5dB/div.
80dB (@ 10dB/div.)
40dB (@ 5dB/div.)
8 x10, logarithmic, dBm
0 - 40dB, 10dB steps
+20dBm (0,1W)
+10dBm
-99.6dBm to+10dBm
–100dBm (9kHz RBW)
better than 75dBc
better than 75dBc
distance>3MHz)
N-Connector
N-Connector
BNC (F)
50Ω , 48MHz,
-30dBm ±2dB
6V (Near field probes)
ATTN ≥10dB
typ. 1.5 : 1
3.5 mm
Center Frequency,
Center Frequency, Reference and
Tracking Generator Level, Marker
Peak Detection
with software AS100E
Mean value measurement
10 set-up settings
Ear Phones
RS-232 control disabled
150kHz to 1050MHz
–50dBm to +1dBm
better than 20dBc
General Data
Temperature Range:
Power Requirements:
Power Consumption
HM 5012-2:
HM 5014-2:
Protective System:
Dimensions (W x H x D):
Cabinet, Weight:
Accessories supplied:
4kHz
Instruction Manual, Power Cord, PC-Software on CDR
Optional Accessories:
Opto-Interface HZ 70, Sniffer Probes HZ 530,
Transient Limiter HZ 560, Line Impedance Stabilization
Network HM 6050-2, VSWR Measuring Bridge HZ 541
±3dB
±1dB
±25 V
Accessories
±1dB
±1dB
50Ω
50Ω
HZ 560 Transient-Limiter
The transient limiter HZ 560 protects a spectrum analyzer input
from damage caused by high level transients from line impedance
stabilization networks during EMC testing for conducted
emissions. The transient limiter can withstand inputs as high as
∅
jack
2W of average power. The built in high pass filter helps reduce 50/
60 Hz line feed through. The use of this limiter is strongly
recommended when measuring with a Line Impedance Stabi-
lization Network.
Specifications
Frequency Range:
Insertion loss:
below 1kHz
over 100MHz
Max. Input level:
DC-voltage
VSWR:
Connectors:
±3dB
Dimensions (W x H x D):
Operating: 10 °C - 40 °C
105 - 250 V~, 50 - 60 Hz
approx. 30 W
approx. 35 W
Safety Class I (IEC 1010-1/EN61010-1)
285 x 125 x 380 mm
Lockable tilt handle, approx.6kg
150kHz to 30MHz
10dB (+1.5 / –0.5dB)
Continuous 2W (+33dBm)
1.5:1 or better
BNC (input and output)
32 x 67 x 32 mm
Subject to change without notice
>90 dB
>50 dB
±50V DC