3.5.2 Audit trail list

Starting from the event list, call up the audit trail list using the
button.
Select current
entry
The following are entered in the audit trail list:
- each operator intervention with time stamp, name, ID, alteration (reason for
alteration). An operator intervention is any alteration of the system resulting
in an alteration of process data, configuration data, or in the audit trail,
- electronic signing,
- configuration alterations,
- fine calibration,
- successful or failed log-in or log-out attempts,
- alteration of user rights and password changes,
- CompactFlash memory card was removed or inserted,
- process data transfer via interface (start, stop),
- manual time change on the recorder or through the setup software,
- time synchronization,
- removing the back panel cover and
- power ON or power OFF.
3 Operation and visualization
Current entry in short form
Call up event list
Current entry in full length