Weighing With Tolerance Range - KERN WTB-N Series Operating Instructions Manual

Table of Contents

Advertisement

Available languages

Available languages

8.4 Weighing with tolerance range

You can set an upper or lower limit when weighing with tolerance range and thus
ensure that the weighed load remains exactly within the set tolerance limits.
Settings:
 In weighing mode press
 Enter upper limit value, the active position blinks.
(example)
 Now enter the lower limit value as described above and
(example)
Audio signal:
The acoustic signal depends on the settings in menu block „F9 BEP".
Options:
• BEP 0 Acoustic signal turned off
• BEP 1 An acoustic signal sounds when load is within tolerance limits
• BEP 2 An acoustic signal sounds when load is beyond tolerance limits
27
display for entering the upper limit appears. The left digit
blinks. A triangle appears above the "+" symbol.
Moreover, the background of the screen turns red.
Mit Correct the numerical value by
decimal using
. If the last digit is entered, confirm input
by
, the upper limit value is hereby saved.
The display changes to the prompt for entering the lower limit
value.
The colour of the display then changes to orange.
confirm input using
entered. The balance changes into weighing mode.
for about 3 seconds, the
, change the
. The lower limit value is thus
WTB-N_WTB-NM-BA-e-1622

Advertisement

Table of Contents
loading

Table of Contents