Fault
Therapy pressure
is not reached.
8

Scope of supply

Variant
Size
S
vented
non-vented
M
vented
non-vented
L
vented
non-vented
Cause
Mask not adjusted
correctly.
Mask cushion
damaged.
Tube system
leaking.
Air escaping at
pressure
measurement
connections.
Nasal mask
WM 26118
WM 26168
WM 26128
WM 26178
WM 26138
WM 26188
Remedy
Readjust mask (see "4.3 Adjust mask
(Figure E)" page29).
Replace mask cushion.
Check push-fit connector and fit of
tube sleeves.
Seal pressure measurement
connections with plug (included in
scope of supply).
JOYCE SilkGel
Scope of supply
Full-face mask
WM 26418
WM 26468
WM 26428
WM 26478
WM 26438
WM 26488
EN