Battery life
Lasts for approx. 200 measurements
depending on the blood pressure and
pump pressure as well as the number
of Bluetooth
Classification
Internal supply, IP21, no AP or APG,
continuous operation, application part
type BF
Data transfer via
The blood pressure monitor uses
Bluetooth
wireless
Bluetooth
®
technology
2402MHz – 2480MHz frequency band,
Transmission power max. 20 dBM,
Compatible with Bluetooth
phones/tablets
The serial number is located on the device or in the battery com-
partment.
Technical information is subject to change without notification
to allow for updates.
• This device complies with European Standard EN60601-1-2
(In accordance with CISPR 11, IEC 61000-4-2, IEC 61000-4-3
and IEC 61000-4-8) and is subject to particular precautions
with regard to electromagnetic compatibility. Please note that
portable and mobile HF communication systems may interfere
with this unit.
• The device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinproduk-
tgesetz) and the standards EN1060-1 (non-invasive sphyg-
momanometers – Part 1: General requirements), EN1060-3
(non-invasive sphygmomanometers – Part 3: Supplementary
connections.
®
low energy technology,
®
4.0 smart-
®
requirements for electro-mechanical blood pressure measuring
systems) and IEC80601-2-30 (Medical electrical equipment –
Part 2 – 30: Particular requirements for the basic safety and
essential performance of automated non-invasive sphygmo-
manometers).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful life.
If using the device for commercial medical purposes, it must
be regularly tested for accuracy by appropriate means. Precise
instructions for checking accuracy may be requested from the
service address.
• We hereby confirm that this product complies with the
European RED Directive 2014/53/EU. The CE Declaration of
Conformity for this product can be found under:
www.beurer.com/web/we-landingpages/de/cedeclarationof-
conformity.php.

12. Replacement parts and wearing parts

Replacement parts and wearing parts are available from the
corresponding listed service address under the stated material
number.
Designation
Universal cuff (22-44 cm) 163.952