8
Appendix A

Optional Parts

Dose Area Product (DAP)
DAP is available as an option DRX-Revolution Nano Mobile X-ray System. The DAP is
configured and calibrated via the monitor display for the use of a DAP measurement
chamber and additional filters. The X-ray tube will support DAP with automatic
population. DAP meter specifications are the following:
Measurement
Upper limit of Response Range
Saturation
Energy Range
Energy Dependence
Chamber Filter Effect
Wireless infrared remote exposure switch
The X-ray tube will support an optional infrared wireless exposure switch. See the User
Guide for Remote Exposure Switches for Mobile X-ray Systems for instructions on using
this tool.
RFID Badge Reader
See DirectView Online Help Topics.
AF0118 | 2018-03-27
Specification
0.8 nC/cGy-cm2
98% at 400 V, 1 Gy/s
40 – 150 kV
5% over the 40–110 kV range
0.20 mm A1 (70 kV)
0.25 mm A1 (70 kV)