Carestream DRX-Revolution Hardware Manual page 52

Nano mobile x-ray system
Hide thumbs Also See for DRX-Revolution:
Table of Contents

Advertisement

Collimator light
beam diaphragm
and visual aids
Computer
Detectors and
rechargeable
batteries
Generator
Tube head assembly
(X-ray tube
head, yoke,
collimator)
X-ray tube and tube
movements
6-2
The collimator LBD source provides an average luminance 160 Lux at 100
cm.
Refer to CFR, Title 21, Volume 8, 1020.31, (d)(2)(ii).
Intel® Core™ Core i5 or better
OS: Windows
Memory: 8 GB minimum
I/O: 1 Ethernet, 2 USB ports
Storage: Can store up to 1700 images
Network: 10/100/1000 Gigabit Ethernet, IEEE 802.11 a/b/g/n Wireless
network
Storage space for one 43 x 43 cm, one 35 x 43 cm, and one 25 x 30 cm
digital flat panel detector. Storage space for one detector battery. The 35
x 43 detector can be coupled with the grid frame assembly when inserted
into bin. For detectors that are ISO 4090 compliant.
5.3 kW @ 100 msec; 7.7 kW @ 13 msec kW X-ray generator
kV range 40 – 110kV
mAs range 0.2 – 20.0 mAs
mA range 30 – 80 mA (not user-selectable)
Movement force
Column rotation (beta)
Tube angulation (alpha)
Holding force
Column rotation (beta)
Tube range 40–110kV
Anode heat capacity 300 Khu
IEC Focal spots: [0.9 – 1.9] equivalent, 0.1 increments.
Target angle = 14 °
Angulation range (alpha)± 110 °
Tilt range (tau) –16 to +98 °(where zero = tube pointing straight down)
When the arm is fully extended, the X-ray tube focal spot is 2123 mm
from the floor
Vertical (zeta - z)
Horizontal (gamma)
Tube tilt (tau)
Vertical (zeta)
Horizontal (gamma)
(alpha)
(tau)
Technical Specifications
53 N
3.5 lbf (15 N)
20 Nm
24 Nm
18 Nm
40 lbf (175 N)
20 lbf (93 N)
4 lbf (17 N)
26 N-m (230 in-lb)
28 N-m (250 in-lb)
AF0118 | 2018-03-27