Operation and Use
Caution
Check for potential obstacles prior to moving the cart using the "inch" feature.
Brakes can be released manually to allow the cart to be moved in cases of lower
battery power. The manual release lever is located directly above the kickplate.
Caution
The manual release for the brakes are to be used only in case of low battery power and then
only to move the cart to a position to be charged where the brakes are to be engaged again. If
the brakes are not functioning for any other reason, discontinue using the cart immediately and
contact your authorized service representative.
The cart is designed to negotiate obstructions up to 50 mm in height. In order to
negotiate an obstruction, tilt the cart back onto its rear wheels by bracing the cart with
your foot on the kickplate while pulling back on the drive handles, and then push the
cart over the obstruction.