Technical Specifications - Carestream DRX-Revolution Hardware Manual

Nano mobile x-ray system
Hide thumbs Also See for DRX-Revolution:
Table of Contents

Advertisement

6

Technical Specifications

Component
Arm Assembly
Batteries
Cart
Collimator
AF0118 | 2018-03-27
Components are listed in alphabetical order.
Rotates ± 239 ° from docked position
The battery life allows for taking up to 40 typical exposures on a new set
of fully-charged batteries.
This assumes fully-charged batteries and exposure technique factors of 70
kV and 10 mAs.
Operates for a minimum of four hours without turning off the main
system computer components (assumes new batteries).
The time to a full charge is typically less than two hours.
When fully-charged, the battery bank produces 72 V (dc) nominal.
There are a total of five lithium iron phosphate (LiFePO4) batteries; each
battery has a voltage of14.4 Vdc and a nominal capacity of 7.5 Ah.
Caution
Damaged Lithium Ion batteries may present a hazard. In the case of an impact,
the system should be disconnected and should not be used until inspected by an
adequately-trained field engineer.
Electronics to track battery health and to recharge the batteries are
included.
Docked speed
Forward – walking speed
Reverse – walking speed
Can stop and hold its position on an incline of 5 °
Caution
Do not park or leave the DRX-Revolution Nano unattended on an incline. The unit
may roll.
Weight 100 kg (220 lb)
Projection radiography range of 40 – 110 kV. Single transverse and
longitudinal crosshairs on the output aperture. Rotates independently of
the X-ray tube
± 120°.
Dose Area Product (DAP) measurement chamber can be fitted, as well as a
paediatric filter.
The slot for the pediatric filter is located in the front of the head, just
below the handle.
Caution
Only use the approved pediatric filter.
Specification
6-1