PowerPac HV – Instruction Manual
Step
Procedure
7.
Select the time mode
DIT TIM
55 W
TIM : UNTIM D
HRS
VHOURS UNTIM D
8.
Start the run
9.
Monitor the run
10.
24
Description
Use the soft keys to select a time mode and the
alphanumeric keypad to enter the run time.
HRS:
5000
V
500
mA
VHOURS: Run time is in units of volt-hours.
50¡
C
UNTIMED: Run is continuous.
Press
to start a run.
The RUN screen is used to monitor run parameters
or to pause a run. The current step, V, voltage
ramping, mA and W, elapsed time versus the
programmed time, and total elapsed time are
displayed on the run screen.
The run may be paused by pressing the PAUSE
soft key or the
To edit the run parameters during PAUSE see step 6.
To Edit preference see Section 3. Edits made
during the run may be saved if SAVE soft key is
pressed at the end of the run. Otherwise the method
will remain as originally created.
To terminate the run, press the
Note: If the temperature reaches 5° over the set
temperature, an alarm will sound intermittently until
the situation is resolved. That is, until the power is
reduced or the temperature of the gel drops. The
alarm sounds similar to the end of run, but repeats.
The Run Completed or Run Terminated screen is
displayed at the end of a run. Press RERUN to
repeat the run or EXIT to display the method step
list.
Run time is in units of hours.
key.
key.