BIO RAD PowerPac HV Series Instruction Manual page 17

Table of Contents

Advertisement

KEY CHIRP
IRDA
ON
CONTRAST
HI
* KEY CHIRP
ON
CHANGE
EXIT
RESET
DATE
ETUP
KEY CHIRP
ON
DATE
12/07/04
* TIME
16:38 HR
CHANGE
EXIT
RE ET
TIME
ETUP
KEY CHIRP
ON
DATE
12/07/04
* TIME
16:38 HR
CHANGE
EXIT
RE ET
TEMP PROBE CK
Temperature Probe Check
PowerPac HV – Instruction Manual
ON
Activates the audible signal for keystroke.
(Default)
OFF
MM/DD/YY
This is the only format available.
Press the change soft key then enter the
correct date using the alphanumeric
keypad.
HH:mm
24 hr format. Factory setting is Pacific
Standard time
Press the change soft key then enter the
correct time using the alphanumeric
keypad.
ON
This preference is only available when
(Default)
the temperature probe is connected.
OFF
ON: The run is paused and error code
18 is displayed in case of failure. A failure
occurs when the temperature remains
5°C less than the set temperature for
more than 15 minutes. This can be caused
by a detached temperature probe or an
insufficient power setting.
OFF: Run will continue regardless the
temperature.