BIO RAD PowerPac HV Series Instruction Manual page 25

Table of Contents

Advertisement

Step
Procedure
8.
Select the time mode
S1/1 EDIT TIME
TIME: UNTIMED
HRS
VHOURS
9.
Add a new step or
save the method
10.
Start the run
11.
Monitor the run
12.
Description
Use the soft keys to select a time mode and the
alphanumeric keypad to enter the run time.
HRS:
500 mA
VHOURS:
400 W
UNTIMED: Run is continuous.
UNTIMED
Note: Steps programmed after an untimed step are
ignored.
Press EDIT to return to the Method Screen.
To program additional steps press the down arrow
key to select the empty step and repeat Steps 5–8.
If the method will not be saved, go directly to Step 10.
Press SAVE and then go to Step 10.
Press DELETE to delete a selected step.
Press the METHOD soft key to return to the Method
List.
Press
to start a run.
The RUN screen is used to monitor run parameters
or to pause a run. The current step, Volts, mA and
Watts values, elapsed time versus the programmed
time, and total elapsed time are displayed on the
RUN screen. The method step is also indicated.
This run is executing Step 1 of 2.
The run may be paused by pressing the PAUSE
soft key or the
To edit the run parameters during PAUSE see
Section 4.5. Edits made during the run may be
saved, if the SAVE soft key is pressed at the end of
the run. Otherwise the method will remain as
originally created.
To terminate the run, press the
The Run Completed or Run Terminated screen is
displayed at the end of a run. Press RERUN to
repeat the run or EXIT to display the method step
list.
PowerPac HV – Instruction Manual
Run time is in units of hours.
Run time is in units of volt-hours.
key.
key.
17