Mounting on pipe socket

In order to ensure a tight seal and correct function of pHix® Compact, the
following should be complied with:
1. The pipe socket must be flushed and free of burrs, etc.
2. The pipe socket must have a length so that the tip of the electrode is
immersed min. 10 mm into the liquid.
3. Take care to have adequate room for mounting and removal of pHix®
Compact.
4. The working pressure of the system must not exceed 5 bar g.
5. The working temperature of the system must not exceed 80° C.
See also the following illustrations for correct adaptation to the pipe system.