MANUAL
EN 4.10 pHix® Compact 1904
pHix® Compact conforms to pressure class PN 5, and can therefore be
mounted in closed systems with a working pressure of max. 5 bar.
Mounting in open channels and reservoirs
Due to the lifetime of the electrode and measurement accuracy, the
following should be complied with when mounting pHix® Compact.
1. pHix® Compact should be mounted on a location with a good liquid
circulation.
2. The electrode tip should be minimum 30 cm below the liquid surface.
3. The electrode tip must not touch the bottom.
4. The cable must not be suspended in a cable bracket.
14