EN 4.10 pHix® Compact 1904
If pHix® Compact is being put into a buffer liquid with the same value as the
switch ring is set to, pHix® Compact will automatically begin to scan for a
buffer with the same value. When the value is found, pHix® Compact will
automatically set the 0-point or span to the correct value, and the LED will
indicate this with two short flashes every 5 seconds.
pHix® Compact is now ready either for measurement or for another buffer
adjustment at another pH value.
pHix® Compact determines automatically if zero point or span needs to be
adjusted depending of the electrode's zero point and the preceding buffer
values.
If pHix® Compact cannot find the pH value for the selected buffer solution, it
will go into alarm mode, which is indicated with continuous LED flashes
every second.
Note! There is only need for calibration of two (buffer-) values, for example,
4-7 pH / 7-10 pH.