Specifications
Electrical Characteristics
Measuring ranges
i3000s (24 and 36 Flex)
i2000 Flex
Output sensitivity
Accuracy (at 25 ºC)
Frequency range
i3000s (24 and 36 Flex)
i2000 Flex
Phase error
Linearity
Position sensitivity
External field
Minimum load
Noise
Adhere to local and national safety codes.
Individual protective equipment must be
used to prevent shock and arc blast injury
where hazardous live conductors are
exposed.
CAT III equipment is designed to protect
against the transients in the equipment in
fixed equipment installations, such as
distribution panels, feeders and short
branch circuits, and the lighting systems in
large buildings.
30 A/300 A/3000 A AC,
switch selectable
20 A/200 A/2000 A AC,
switch selectable
100 mV/10 mV/1 mV per A
(AC coupled)
±1 % of range (45 – 65 Hz)
10 Hz to 50 kHz (-3 dB)
10 Hz to 20 kHz (-1 dB)
< ±1° (45 – 65 Hz),
±10° (at 20 kHz)
±0.2 % of reading from
10 %
- 100 % of range
±2 % of range with cable
>25 mm (1") from the
coupling
±1 % of range with cable
>200 mm (8 in) from the
head
100 ke for specified
accuracy
8 mV RMS (0.3 % of
range) in 30 A range
2 mV RMS (0.1 % of
range) in 300 A and
3000 A