General Data; Input And Output Quantities - Omicron C-Probe 1 User Manual

Ac/dc current clamp
Table of Contents

Advertisement

AC/DC Current Clamp C-Probe 1
3.2
3.3
Table 3-1:
Accuracy data
10

General Data

Dimensions:
Max. wire diameter:
Max. opening of clamp jaws:
Connecting cable:
Weight:

Input and Output Quantities

Maximum system voltage of live wire: V
Bandwith of frequency: DC up to 10 kHz.
Switch position
100 mV/A
Measurement
1
range
Accuracy
10-100Hz
meas. value
Accuracy
DC
meas. value
Accuracy
10 Hz - 1 kHz
meas. value
Accuracy
1.5/-7% of
10 Hz - 10 kHz
meas. value
Phase shift
45 - 55Hz
Phase shift
10 Hz - 1 kHz
Phase shift
10 Hz - 10 kHz
1
A higher modulation leads to the saturation of the core and hence to faulty measurements.
This does not affect safety.
230 x 67 x 36 mm
11.8 mm
15 mm
2 m length, isolated banana plugs
330 g (battery included)
= 600 V with reference to ground
eff
0-10 A
0-40 A
AC/DC
AC/DC
1% of
2% of
meas. value
± 5 mA
± 30 mA
1,5% of
2% of
meas. value
± 5 mA
± 30 mA
1,5% of
2% of
meas. value
± 5 mA
± 30 mA
0/-8% of
meas. value
± 5 mA
± 30 mA
< 0.5°
< 0.5°
< 2°
< 2°
< 30°
< 30°
10 mV/A
40-60 A
60-80 A
AC/DC
AC/DC
0/-8% of
0/-12% of
meas. value
meas. value
± 30 mA
± 30 mA
0/-10% of
0/-10% of
meas. value
meas. value
± 30 mA
± 30 mA
0/-10% of
0/-12% of
meas. value
meas. value
± 30 mA
± 30 mA
0/-10% of
0/-12% of
meas. value
meas. value
± 30 mA
± 30 mA
< 0.5°
< 0.5°
< 3°
< 3°
< 30°
< 30°

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents