Typical Noise Levels At Output (Rms); Influencing Factors; Voltage Supply - Omicron C-Probe 1 User Manual

Ac/dc current clamp
Table of Contents

Advertisement

3.4
Table 3-2:
Typical noise levels
3.5
3.6
The above mentioned measurement deviations relate to the following data:
ambient temperature of 18 °C to 28 °C
relative humidity between 20 % and 75 %
wire centred and parallel to the marking on the clamp
sinus signal form for AC measurement
impedance of the connected measuring device >100 k Ohm
battery voltage 8 V
DC
the operational condition of the clamp one minute after it was switched
on.

Typical Noise Levels at Output (RMS)

Frequency band
Range 10 mV/A
Range 100 mV/A

Influencing Factors

Position of wire in the clamp:
Temperature variations:

Voltage Supply

Battery:
Battery operating period:
Current consumption:
Control indicator:
± 0.5 V
DC up to 1 Hz
μ
15
V
1.5 mV
Max. ± 0.5 % of displayed value.
Wire from stop to stop.
800 ppm/°C max. within range of
±
-10 °C up to +50 °C.
9 V Alkali block battery, type 6 LR 61
70 h typically, at least 40 h
6 mA typically, 12 mA max.
Green LED, off at a battery voltage
< 6,5 V.
Technical Data
1 Hz up to 10 kHz
μ
140
V
14 mV

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents