Product specifications

Physical
characteristics
Operating
conditions
Technical
specifications
8
Instrument type:
Instrument dimensions:
Weight:
Operating power:
Frequency:
Electrical input:
IMPORTANT! If the supplied power fluctuates ±10% beyond the rated
voltage, a power line regulator may be required. High or low voltages can
adversely affect the electronic components of the instrument.
Installation site:
Altitude:
Operating temperature:
Operating humidity:
Pollution degree:
Dynamic range:
Processing time:
Light sources:
Excitation filters:
Emission filters:
Detectors:
Calibration type:
Sample chamber:
Tube type:
Warm-up time:
Benchtop fluorometer
7.3 in (w) × 11.1 in (l) × 4.1 in (h)
(18.6 cm × 28.2 cm × 10.3 cm); rectangular shape
60 oz. (1.7 kg)
100–240 ±10% VAC, 1.3 A
50/60 Hz
48 VDC, 1.87 A
Indoor use only
Between sea level and 2000 m (6500 ft.) above
sea level
10–30°C
15–80% (non-condensing)
The instrument has a Pollution Degree rating of II.
The instrument may only be installed in an
environment that has nonconductive pollutants.
Typical environment with a Pollution Degree II
rating are laboratories and sales and commercial
areas.
4 orders of magnitude
≤3 seconds/sample
Blue LED (max 460–480 nm)
Red LED (max 620–640 nm)
Blue 456–484 nm
Red 612–644 nm
Green 513–563 nm
Far-Red 671–693 nm
Photodiodes; measurement capability from
320–1100 nm
2- or 3-point standard
Accommodates one Qubit
Qubit
Flex Tube Strip (8× 0.2-mL thin-wall
polypropylene tubes; Cat. No. Q33252)
<35 seconds
Qubit
Flex Tube Strip
Flex Fluorometer User Guide