Summary of Contents for Hydac FluidAqua Mobil Series
Page 1
FAM 10/15 Series FluidAqua Mobil + AquaSensor AS1000 + ContaminationSensor CS 1000 Operating and Maintenance Instructions English (translation of German original) Dokumentation no.: 3417360...
HYDAC Filtertechnik GmbH all rights reserved All rights reserved. This manual may not be reproduced in part or whole without the explicit written agreement from HYDAC Filtertechnik. Contraventions are liable to compensation. Exclusion of liability We made every endeavor to ensure the accuracy of the contents of this document.
Transfer by pumping (dewatering, filtration and degassing) ........... 23 FAM Set-up and Connection ....................24 Setting up the FAM ......................24 Hydraulic Connection of the FAM ..................24 Connecting the suction port connection (IN) ..............24 HYDAC Filtertechnik GmbH Page 3 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
There may be deviations in technical details, figures, and dimensions as a result. If you discover errors while reading the documentation or have additional comments or suggestions, contact us at: HYDAC Filtertechnik GmbH Division Service technology / Filter systems Technical Documentation Department Postfach 12 51...
After modification or repair work that affects the safety of the product has been car- ried out on components, the product may not be returned to operation until it has been checked and released by a HYDAC technician. Please notify us immediately of any modifications made to the product whether by you or a third party.
The index is incremented every time the manual is revised or changed. Safety Information and Instructions These operating instructions contain the key instructions for properly and safely op- erating the FluidAqua Mobil. HYDAC Filtertechnik GmbH Page 8 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
DANGER denotes situations which can lead to death if DANGER safety precautions are not observed. WARNING denotes situa- WARNING tions which can lead to death if safety precautions HYDAC Filtertechnik GmbH Page 9 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Before any maintenance or repair work is carried out on the unit, electrical power to the unit must be disconnected, and all hydraulic pressures relieved. HYDAC Filtertechnik GmbH Page 10 / 80 BEWA FAM10 3417360 en.doc...
(e.g. warning sign or padlock on the main switch). When the product is supplied in partial consignments, the safety devices are to be applied by the operator as specified by law/pertinent regulations. HYDAC Filtertechnik GmbH Page 11 / 80 BEWA FAM10 3417360 en.doc...
(visual check once a month). Any loose connections or damaged cables are to be replaced immediately. DANGER Hydraulic systems are under pressure The hydraulic system must be depressurized before performing any work HYDAC Filtertechnik GmbH Page 12 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Do not operate the FAM unless all the safety devices function properly. The product is to be checked once a day for external damage and the proper func- tioning of the safety devices. HYDAC Filtertechnik GmbH Page 13 / 80 BEWA FAM10 3417360 en.doc...
Do not make any modifications (design modifications, extensions) to the FluidAqua Mobil without the prior consent of the manufacturer. Any design modifications or extensions may not be made without HYDAC Fil- tertechnik GmbH’s express prior written approval. Immediately replace any machine components which are not in perfect condition.
The unit must be evacuated completely and the inlets and outlets must be closed before transport. Suction and pressure hoses and connection cables are coiled on the holders pro- vided for that purpose and fastened in place. HYDAC Filtertechnik GmbH Page 15 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Train/truck During transport, the FAMs that have castors must be placed that on supports that are high enough to ensure that there is no load on the castors. HYDAC Filtertechnik GmbH Page 16 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
The fluid filter that is installed provides efficient filtering-out of solid materials. The degassing of the medium is achieved by the vacuum set up in the vacuum con- tainer. HYDAC Filtertechnik GmbH Page 18 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
The FluidAqua Mobil can only be utilised in connection with the following hydraulic fluids: Mineral oils acc. DIN 50524 Gear oils acc. DIN 51517 and DIN 51524 Synthetic esters (HEES) acc. to DIN 51524/2 HYDAC Filtertechnik GmbH Page 21 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
• the operating status can be queried • the FAM can be switched on/off by a level sensor in the transfer by pumping operation mode. HYDAC Filtertechnik GmbH Page 22 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
The filling level in the clean oil tank must be monitored permanently during this op- erating mode in order to exclude the possibility of overfilling. HYDAC Filtertechnik GmbH Page 23 / 80 BEWA FAM10 3417360 en.doc...
Connecting the return port connection (OUT) The return-flow line or the return-flow hose must be installed below the level of the oil in order to prevent air from entering the medium. HYDAC Filtertechnik GmbH Page 24 / 80 BEWA FAM10 3417360 en.doc...
Synthetic vacuum pump oil in accordance with DIN 51506, Group VDL, ISO VG100 must be used for topping up. This can be ordered from HYDAC under the part. no.: Description Part no.: Vacuum pump oil VE101...
(The working temperature is preset to 55°C; if 80°C is exceeded, then the safety thermostat will shut the device off.) Fluid temperature 55°C Safety thermostat 80°C HYDAC Filtertechnik GmbH Page 26 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
50 ... 100 mbar (e.g. from 300 mbar to 400 mbar). The FAM will not start up again automatically when voltage is restored after a volt- age failure; it must be restarted instead. HYDAC Filtertechnik GmbH Page 28 / 80 BEWA FAM10 3417360 en.doc...
• If the main screen from Section 0 appears on the display, then the after-running phase has been completed and the FAM is ready for operation again. • Now the FAM can be switched off via the main switch. HYDAC Filtertechnik GmbH Page 29 / 80 BEWA FAM10 3417360 en.doc...
Is shown during the starting phase and during automatic operation. RESET Is displayed only when there is a fault The most recently selected operating mode will always be shown in the start screen. HYDAC Filtertechnik GmbH Page 31 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
OPERA- SERVICE MENU1 F1 = OPERATOR F2 = SERVICE F4 = MENU1 OPERATOR Manual operation for operator SERVICE Manual operation for HYDAC Service (password-protected) MENU1 Back to MENU1 HYDAC Filtertechnik GmbH Page 34 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Language selected: ENGLISH SELECT LAN- MENU1 GUAGE F1 = Select language F4 = back to "MENU1" SELECT Press the F1 key repeatedly until the desired language ap- LANGUAGE pears. HYDAC Filtertechnik GmbH Page 36 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Rel. humidity: Actual: Display of the current moisture at the FAM inlet. A display of 0 ... 100% is possible. Temp. Medium temperature at inlet to the FAM. HYDAC Filtertechnik GmbH Page 37 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Is displayed only in the presence of faults Display: ISO: Nominal: The nominal value can be changed from 0 ... 25% (see chapter 0). ISO: Actual: Display of the current contamination at the FAM inlet. HYDAC Filtertechnik GmbH Page 38 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
If a water saturation level of 90% occurs during automatic operation, then the oper- ating mode switches automatically in accordance with the following table: Selected operating mode: changes automatically after: => EW+FTB => DW+FT HYDAC Filtertechnik GmbH Page 40 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
- depending on the project planning. Calling up the Edit mode – redirects the entry via ends the Edit mode and confirms the entry Navigating in the list HYDAC Filtertechnik GmbH Page 41 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
The expanded character set is available in fields that have the "String" format. Date and Time Values for date and time are entered in accordance with the alphanumerical values. HYDAC Filtertechnik GmbH Page 42 / 80 BEWA FAM10 3417360 en.doc...
The following approaches are possible for entering numerical and alphanumerical characters: • Entering a value, which causes the existing value to be replaced completely • Modifying the character of an existing value HYDAC Filtertechnik GmbH Page 43 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
All of the upper and lower case letters will be displayed first, and afterwards the special characters and numerals. Press when characters are to be entered. First a few special characters are displayed, and then the numerals. HYDAC Filtertechnik GmbH Page 44 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Page 45
You can begin the entry at any given position in the entry field. The algebraic sign can also be modified during the entry at any position in the entry field SHIFT with HYDAC Filtertechnik GmbH Page 45 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
4. If the password is complete, activate the "OK" button with press . You can abort the process with "Cancel". After logging in, you can operate the operating objects that have password protec- tion. HYDAC Filtertechnik GmbH Page 46 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Page 47
Password The password and the log-out time are encoded in a list and stored on the operating device in a manner that is not affected by network power outages. HYDAC Filtertechnik GmbH Page 47 / 80 BEWA FAM10 3417360 en.doc...
One can used the arrow keys to scroll through the individual fault messages. In order to restart the device, first the error must be eliminated and then acknowl- edged afterwards. It is only then that automatic operation is possible again. HYDAC Filtertechnik GmbH Page 48 / 80 BEWA FAM10 3417360 en.doc...
Page 49
Maintenance Instructions for the Instructions for the ContaminationSensor. ContaminationSensor. Fault, CS device error For details, see Operat- For details, see Operat- ing and Maintenance ing and Maintenance HYDAC Filtertechnik GmbH Page 49 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Page 50
Vacuum chamber over- Emergency Off Max Check pressure in vac- HYDAC Filtertechnik GmbH Page 50 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Page 51
400 mbar ab- soute) FAM tray too full. Drain FAM oil pan too full Drain oil pan, check the and seal leaky site of FAM for leaks FAM. HYDAC Filtertechnik GmbH Page 51 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
3 5 minutes The pressure in the vacuum Evacuation pump not column is too low conveying Raise pressure (e.g. check from 300 mbar to 400 mbar). HYDAC Filtertechnik GmbH Page 52 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Carry out check for leakage and eliminate leaks if present. Restart FAM with the F1 key. Fault appears again Check float switch and sup- Failure repaired ply lines HYDAC Filtertechnik GmbH Page 53 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
If an engine is replaced at the same time, then the direction of rotation of the engine in manual operation must be checked before the FAM is switched On. HYDAC Filtertechnik GmbH Page 57 / 80 BEWA FAM10 3417360 en.doc...
Disconnect plug at level sensor. Remove level sensor Manually check the functioning of the switching points Check cable to and in the Replace level sensor, control cabinet because defective Visual check in the vacuum column HYDAC Filtertechnik GmbH Page 58 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Replace level sensor, because defective Plug + Check cable to the control cabi- and in the control cabinet to the SPS. Functioning? Replace dam- SPS defec- aged parts tive HYDAC Filtertechnik GmbH Page 59 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Actuate ESC key and reset malfunction with the F3 key. Restart FAM with the F1 key. Fault appears again Pull out mains plug, check Failure repaired clogging indicator and its plug, replacing as neces- sary HYDAC Filtertechnik GmbH Page 60 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Check the viscosity of the medium (only with motor pro- tection switch) Clean the tank of the water ring vacuum pump if neces- HYDAC Filtertechnik GmbH Page 61 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Page 62
Fault appears again? Replace motor protection switch and/or fuse Close switch cabinet door, main switch to "On", start FAM Fault appears again? Contact the HYDAC Ser- vice Department HYDAC Filtertechnik GmbH Page 62 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Replace the sensor and/or have it checked and then reinstall it Reconnect mains switch, main switch to "On", start device with "F1" key Error appears again SPS defective, Failure repaired Contact the HYDAC Service Department HYDAC Filtertechnik GmbH Page 63 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Page 64
FluidAqua Mobil – FAM 10/15 +CS1000 +AS1000 Error messages Replace cable and/or plug Reconnect mains switch, main switch to "On", start de- vice with "F1" key Error appears again Failure repaired HYDAC Filtertechnik GmbH Page 64 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
The secondary side of the transformer (control current circuit) has 1A 250V The secondary side of the transformer (control current circuit) has 1A 250V Contamination Sensor Display CSD (Optional) Contamination Sensor CS (Optional) HYDAC Filtertechnik GmbH Page 67 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
After the fault has been corrected, the safety thermostat can be reset by pressing the green key in the head. To protect yourself from burns, wait until the heater has cooled down before doing so. Reset button HYDAC Filtertechnik GmbH Page 68 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
The oil may need to be changed sooner than that in the event of heavy contamination. After 2000 hours of operation or if oil mist escapes from the separator during the operation of the pump the exhaust filter must be changed. HYDAC Filtertechnik GmbH Page 69 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Inlet flange Rotor Slide Exhaust valve Oil sump Exhaust cover Air de-oiling element Oil filling screw plug Gas ballast valve The following descriptions are in reference to these positions. HYDAC Filtertechnik GmbH Page 70 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
• Replace ventilation cover seal (10). • Assembly in reversed order of sequence. For further information please refer to the attached operating instruction for the BUSCH vacuum pump. HYDAC Filtertechnik GmbH Page 71 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...
Moisten the O-rings on the new filter element with the operating medium. Now the new filter element can be installed. Make sure the new element is fitted tightly to the connection part. HYDAC Filtertechnik GmbH Page 72 / 80 BEWA FAM10 3417360 en.doc...
After disassembling the unit and separating the various materials, they can be re- used or disposed of properly in accordance with local regulations. HYDAC Filtertechnik GmbH Page 73 / 80 BEWA FAM10 3417360 en.doc...
FluidAqua Mobil – FAM 10/15 +CS1000 +AS1000 Technical data Type label The type plate is located on the front panel. Please add the data from the type plate into the following copy. HYDAC Filtertechnik GmbH Page 75 / 80 BEWA FAM10 3417360 en.doc 2008-06-06...