CALCulate[1]|2|3|4:LIMit:LOWer[:DATA]

Syntax
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA] <numeric_value>
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA]? [MIN|MAX]
Description
This command enters a value for the lower test limit for the specified CALCulate block
used in the CALCulate[1]|2|3|4:LIMit:FAIL? test. The units used are dependent
on the current setting of UNIT:POWer and CALCulate:RELative:STATe as shown in
Table 4-1.
When the measured value is less than the value specified in
CALCulate[1]|2|3|4:LIMit:LOWer[:DATA],
CALCulate[1]|2|3|4:LIMit:FAIL? reports a fail. When the measured value is
greater than or equal to the limit, a fail is not reported.
The query returns the current setting of the lower limit or the values associated with MIN
and MAX for the specified CALCulate block.
Table 4-1
Measurement mode
Single channel
Ratio
Difference
Parameter
Item
numeric_value
Keysight U2040 X-Series Programming Guide
Measurement units
Measurement type
Avg, Pk
Pk–Avg
Avg, Pk, Pk–Avg
Avg, Pk
Pk–Avg
Range of values
–150 to +230 dBm or
–180 to +200 dB
DEF
MIN
MAX
CALCulate Subsystem
CALC:REL:STAT OFF
CALC:REL:STAT ON
Linear
Log
Linear
Watt
dBm
%
%
dB
%
%
dB
%
Watt
dBm
%
%
dB
%
Description/Defaul t value
A numeric value for the lower test limit:
– DEF: the default is –90 dBm or –120 dB
– MIN: –150 dBm or –180 dB
– MAX: +230 dBm or +200 dB