7. Unpacking the System

As you unpack your Nupro Freedom™ Cordless Prophy System with SmartMode™ Technology, verify that the following
components and accessories are included:
System:
(1) Cordless Handpiece Inner Module
(1) Outer Sheath
(1) Charging Base for Cordless Handpiece
(1) Power Supply and Cord
(20) Nupro Freedom
Disposable Prophy Angles (DPAs)
®
(1) Disposa-Shield
Disposable Barrier Trial Pack (25)
®
(1) Cradle for Cordless Handpiece
Package:
(1) Cordless Handpiece Inner Module
(3) Outer Sheaths
(1) Charging Base for Cordless Handpiece
(1) Power Supply and Cord
(200) Nupro Freedom
®
(1) Box of Disposa-Shield
(1) Cradle for Cordless Handpiece

8. Infection Control Procedures

The objective of the information provided in this section is to reduce the potential for patient cross contamination when
using a Nupro Freedom™ Cordless Prophy System with SmartMode™ Technology during routine dental care. In the event
any regulatory agency disagrees with this information, the agency requirements take precedence.
NOTE: Outer sheaths must be steam-autoclave sterilized prior to each use.
Additional outer sheaths are available for purchase.
Instructions for Steam Sterilizing the Outer Sheath
WARNINGS
LIMITATIONS ON
REPROCESSING
POINT OF USE
CONTAINMENT AND
TRANSPORTATION
PREPARATION FOR
DECONTAMINATION
English • 6
Disposable Prophy Angles (DPAs)
®
Disposable Barriers (500 Barriers)
These instructions are for use ONLY on the outer sheath. All other parts of the sys-
tem should be disinfected according to the procedures in the "Disinfecting All Other
Parts" section.
The outer sheath for the Nupro Freedom™ Cordless Prophy System with
SmartMode™ Technology is not sterile upon receipt and must be steam-autoclave
sterilized prior to use in accordance with the following instructions.
Do not use an automatic washer/disinfector for the outer sheath. Discoloration of mate-
rial will occur.
Repeated cleaning and steam sterilization cycles have minimum effect on these
instruments. End of life is normally determined by wear and damage due to use.
Do not use chemical disinfectants prior to sterilization or rapid deterioration of the
material may occur.
Cold liquid disinfection/sterilization, chemical vapor sterilization, and dry heat ster-
ilization methods have not been tested or validated for efficacy and are not recom-
mended for use.
Do not immerse the outer sheath in an ultrasonic bath.
Remove excess soil with disposable cloth or paper wipe.
It is recommended that instruments are reprocessed as soon as is reasonably practi-
cal following use.
Protect the outer sheath from contact with other dental instruments that may cause
damage.
Remove the outer sheath from the cordless handpiece inner module of the cordless
handpiece. Only the outer sheath may be steam-autoclave sterilized.