Start Test; Stop - Interacoustics EyeSeeCam Instructions For Use Manual

Hide thumbs Also See for EyeSeeCam:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
The correct plane for LARPs
Figure 4.5.5-2 Direction/Plane guide
The last impulse performed is always shown with the darker black line. If your impulses are in the correct
direction/plane you will see your response occur in the shaded areas. The blue shaded areas are for the
LA or LP and the red shaded areas are for the RA or RP.

Start test

Select Start and begin performing RALP or LARP impulses.
Figure 4.5.6-1 Data collection screen with an accepted Right Anterior Impulse

Stop

After a reasonable number of impulses are collected in both planes (7-10 anterior and 7-10 posterior) you
can hit Stop which will end the data collection. Or you can use the auto stop as described in section
4.4.5.
EyeSeeCam vHIT - Instructions for Use - EN
The correct plane for RALPs
Page
34

Advertisement

Table of Contents
loading

Table of Contents