Pipe Mounting; Wall Mounting; Multiple Wet-Section Setup - ABB Navigator 500 Series Operating Instructions Manual

Sodium analyzer
Hide thumbs Also See for Navigator 500 Series:
Table of Contents

Advertisement

Navigator 500
Sodium analyzer

3.2.4 Pipe mounting

Dimensions in mm (in.)
194
(7.64)
Pipe diameters:
max. 62 (2.44) / min. 45 (1.77)
Vertical
configuration
Fig. 3.8 Transmitter pipe-mount options
OI/ASO550–EN Rev. G
213 (8.38)
Cable gland kit
– optional
98 (3.85)
66 (2.6)
Pipe-mount kit
– optional
Horizontal
configuration

3.2.5 Wall mounting

Dimensions in mm (in.)
214 (8.42)
62
(2.44)
66
(2.40)
Fig. 3.9 Transmitter wall-mount option

3.3 Multiple wet-section setup

Fig. 3.10 shows the multiple wet-section setup (4 wet-sections
maximum).
Note.
Maximum length of cable from transmitter to
single-stream wet-section(s) = 30 m (92 ft.).
Multi-stream wet-sections cannot be connected.
Transmitter
Maximum length of cable –
transmitter to (last) single-stream
L2
L1
Multiple wet-sections connected in series
L1 + L2 + L3 + L4 = <30 m (92 ft.)
Fig. 3.10 Multiple wet-section setup – maximum cable length
3 Installation
200 (7.87) fixing centers
Ø6.4
(0.25)
Cable gland kit
– optional
30 m (92 ft.)
wet-section
L3
L4
13