5.3 Performing a calibration for the first time –
multi-stream wet-section
To perform a quick calibration from an Operator page:
1. Press the
key (below the Cal prompt).
Navigator
Tag 1
09:57 30/01
2.93
Tag 2
09:57 30/01
2.75
Tag 3
09:57 30/01
2.25
The Calibrate page is displayed:
Menu
Calibrate
Exit
2. Press the
key (below the Select prompt).
The Calibrate / Sodium Three Stream page is displayed.
Calibrate
Sodium Three Stream
Back
3. Press the
key (below the Select prompt on the
Calibrate / Sodium Three Stream page).
The Sodium Three Stream / Start Calibration page is
displayed:
Sodium Three Stream
Start Calibration
Exit
OI/ASO550–EN Rev. G
10 : 34: 52
2012–04–15
ppb
ppb
ppb
CAL
Select
Select
Yes
4. Press the
key (below the Yes prompt on the Sodium
Three Stream / Start Calibration page).
The Calibration page is displayed with a bar graph to
indicate calibration progress:
Sodium Three Stream
PV 2.93
SS1 100
Please Wait
Abort
At the end of the calibration, and if the calibration is
successful, the Active Slope, Active Offset, Last Slope and
Last Offset values in the Signals View are updated – see
Section 6.6, page 33 for details of View modes.
Note. The Active Slope and Active Offset values are
not updated if the calibration fails. The Calibration Log
is updated with the Active Slope and Active Offset
values.
5. To exit the Calibration page, press the
OK prompt). Calibration continues and calibration
progress can be monitored from the Calibration View –
see Fig. 5.1, page 24.
The Calibrate / Sodium Three Stream page is displayed.
Calibrate
Sodium Three Stream
Back
6. Press the
key (below the Back prompt on the
Calibrate / Sodium Three Stream page).
The Calibrate page is displayed:
Menu
Calibrate
Exit
7. Press the
key (below the Exit prompt) to return to the
Operator page.
5 Calibration
ppb
ppb
OK
key (below the
Select
Select