ABB Navigator 500 Series Operating Instructions Manual page 47

Sodium analyzer
Hide thumbs Also See for Navigator 500 Series:
Table of Contents

Advertisement

Navigator 500
Sodium analyzer
Menu
Calibration Setup
Low Std Solution
High Std Solution
Recovery Time
Grab Sample Time
Restore Defaults
Hold Outputs
OI/ASO550–EN Rev. G
Comment
Sets the low standard calibration solution value between 50 and
10,000 ppb.
Sets the high standard calibration solution value between 50 and
10,000 ppb.
Sets the recovery time in minutes (between 10 and 60 minutes).
This value can be adjusted to ensure the wet-section has adequate
time to reach a stable reading after a calibration.
The default setting is 15 minutes, but this may need to be extended if
the process concentration is very low; especially if a high calibration
has been carried out.
Sets the grab sampling time in minutes.
Restores Calibrate parameters to their default values / settings.
Enables / Disables the Hold Outputs function.
Outputs include current outputs and alarms.
7 Menu descriptions
Defaults
100 ppb
1000 ppb
15 min.
15 min.
45