ABB Navigator 500 Series Operating Instructions Manual page 49

Sodium analyzer
Hide thumbs Also See for Navigator 500 Series:
Table of Contents

Advertisement

Navigator 500
Sodium analyzer
Menu
Solution Detection
Flow Meter
Flow Meter 1 – 3
Stream Settings
Sample Time
Sequence
Stream 1 (3)
Man. Valve Control
Low Standard Valve
High Standard Valve Enables the High Standard Valve for manual control.
Stream Valve
Stream Valve 1 – 3
Regeneration Valve
Reset to Defaults
OI/ASO550–EN Rev. G
Comment
Enables / disables the solution detection function.
Note. Single-stream wet-sections only.
Enables / disables flow measurement.
Note. Multi-stream wet-sections only.
Enables / disables flow measurement for the selected stream.
Note. Multi-stream wet-sections only.
Sets the how long each stream is sampled.
Sets the order in which the streams are sampled, a maximum of 8
entries are permitted. A 0 indicates the end of the sequence.
An example sequence for streams 1 to 3 is: 1,2,3,1,2,3,0 where the 3
streams are sampled consecutively.
Enables / Disables the selected stream (disabled streams are not sampled).
The solenoid valve can be energized manually. The current
concentration value is shown for reference. When the Man. Valve
Control parameter is exited, the valve returns to the closed state.
Note: Press the
key to open the valve manually, press the
to close the valve manually.
Enables the Low Standard Valve for manual control.
Note. Single-stream wet-sections only.
Enables the Stream Valve for manual control.
Note. Multi-stream wet-sections only.
Enables the selected Stream Valve for manual control.
If fitted, enables the Regeneration Valve for manual control.
Resets all wet-section parameters to their default values.
7 Menu descriptions
Defaults
Disabled
Disabled
15 min.
1,2,3,1,2,3,1,2,1
(3-stream)
key
Disabled
47