Remote Output Signals - Cablescan 128/PC Instruction Manual

Continuity tester and fixture
Table of Contents

Advertisement

Page 36

REMOTE OUTPUT SIGNALS

Two output signals are provided for controlling external devices through unused pins on the
printer/serial connector. The pins are connected to the CPU board where jumpers may be
installed to activate the signals. Refer to Figure 13 for jumper locations. The output signals are:
Pass/Fail and End of Test, and are brought out on pins 24 and 25, respectively. A jumper also is
provided for supplying +5vdc at pin 23. The pin assignments and labels used on the CPU board
are as shown below:
Function
Pass/Fail
End-of-Test
Ground
+5vdc
The Standard Jumper settings, using W1 and W2, match the jumper
locations for the Model 256 and Model 256/PC testers. The Alternate
Jumper Setting, using W4 and W5 in lieu of W1 and W2, outputs the
signals on the same connector pins as the Model 512 jumper settings.
Refer to Figure 13 for jumper locations on the CPU board.
END-OF-TEST
PASS/FAIL
(Passed)
PASS/FAIL
(Failed)
Standard Jumper
Setting
W2
W1
-
W3
2 ms
Timing is approximate and can vary between testers.
Alternate Jumper
Setting
Pin 24
W4
Pin 25
W5
Pin 7
-
Pin 23
W3
NOTE
15 ms
NOTE
Model 128/PC Instruction Manual
Pin 25
Pin 24
Pin 7
Pin 23
2 ms