VP-ITC MicroCalorimeter User's Manual
Inserting the new syringe plunger tip
To insert the new tip:
The Guide Sleeve and Adjustment Collar must be in place as described on the previous
page.
Select the Install New Tip button
Place a new plunger tip (open end of plunger tip facing out) into the end of the Tip Pusher
Tool.
Insert the Tip Pusher Tool through the center of the Guide Sleeve and over the end of the
plunger.
Pull out the Tip Pusher Tool to remove the Plunger Tip.
Leave the Threaded Cap and Guide Sleeve on the Pipette Injector to prepare for tip polishing.