Section 5: Instrument Maintenance

The maintenance requirements of the VP-ITC are simple, but essential for quality experiments.
The cells, injection and filling syringes must be kept clean.
The Plunger Tips for the Pipette Injection Syringe must be replaced regularly to prevent
excess wear that would allow the titrant to leak from the syringe.
Y-axis calibration is usually not a problem, but it is good practice to confirm the
calibration every 3 months.
The Injection Syringe needle must be straight.
The Plunger Screw must be kept straight.
The syringe height (i.e. position of the stir paddle from the bottom of the cell) should be
approximately 3 mm. Once the syringe height adjustment is set, it is not usually a
problem, but this distance is custom for each instrument. This means that the user must
set the height position of newly purchased replacement syringes.
This section is arranged as follows:
5.1 Cleaning The Cells and Pipette ...........................................................................................................65
5.2 Replacing Pipette Plunger Tips ...........................................................................................................68
5.3 Y-axis Calibration ...............................................................................................................................72
5.4 Replacing Injection Syringes & Height Adjustment ...........................................................................74
............................................................................................64
64

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VP-ITC and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF