Pavone Systems MCT 1302 Technical Manual page 45

Touch screen batching version with rs232 serial, analog and fieldbus
Table of Contents

Advertisement

FULL SCALE [1301-1302]
Programming of the capacity (net) of the weighing system.
Values: from 0 to 999999
Default: 0
FIXED TARE OF WEIGHING SYSTEM [1106-1107]
Programming of the fixed tare value of the weighing system.
Values: from 0 to 999999
Default: 0
CALIBRATION TYPE
Selection of the type of calibration. Upon confirmation, one of the following procedures is started.
DEAD WEIGHT CALIBRATION [501 ÷ 503]
Calibration of Zero and Full Scale up to 5 points of linearization with the use of sample weights.
TABLE CALIBRATION
[1151 ÷ 1172]
It allows manual programming of up to 5 calibration points. The values correspond to those determined
by the linearization procedure with sample weights. In this way, it is possible to copy the calibration
values made with sample weights.
Page 43