11-3 Axial 
When the vibration gauge measurement value is the axial direction, it must be placed on
the level of the rotation axis, do not put too high, so as not to increase the vibration value.
Put the higher position, the greater the value of the horizontal vibration. At this
time, the Z axis of the vibration gauge is the horizontal direction. If you measure only
one point, it is recommended to use this point. Vibration measurement of the vertical direction,
to be placed on top of the rotating shaft, then the vibration of the Z-axis is the vertical
direction, please note that the two legs of the magnet parallel to the axis of rotation, it can
be fixed in the motor arc on. 
 
 
 
 
 
AG-300 Plus Gateway User Manual
 
18
Page 18
 
Y
 
X
 
Y
 
X
 
 
Z
X
 
 
X
 
When using the axial installation
position, the figure represents the
measured correct vibration axis.
 
Z
 
X
 
 
X
 
 