Ident profile, Add-on instruction for Rockwell systems
Function Manual, 01/2019, C79000-G8976-C410-03
RFID system Meaning
RF200,
This byte is bit-coded. As default, it has the value "B#16#0". With this byte, special
RF300
controllers can be implemented on the CM/reader.
Bit
Description
1
0x01 = Flashing of the "ERR" LED is reset by a "WRITE-CONFIG".
With RF200/RF300, this resets the flashing of the "ERR" LED on the
communications module and on the reader.
RF600
This byte is bit-coded. As default, it has the value "B#16#0".
Bit
Description
1
0x01 = Flashing of the "ERR" LED of the communications module is
reset by a "WRITE-CONFIG".
2
RF200
0x00 (reserved)
RF300
Note: This parameter is intended for trained users. Siemens recommends that
(only
untrained users use the default value.
RF380R)
Readers of the 1st generation: With this parameter you can change the transmit
power (output power) of the RF380R reader of the 1st generation (6GT2801-
3AB10). When doing this, remember that the change to the transmit power will
affect the detection in the limit range (upper/lower operating distance), as well as
the minimum distance that is to be maintained between adjacent RF380Rs.
Settings outside the specified range have the effect that the default value (1.25 W)
will be set. In this case, for reasons of compatibility, there is no error message.
Readers of the 2nd generation: This setting is not necessary with the RF380R
readers of the 2nd generation (6GT2801-3BAx0) because the power limits are
optimized automatically depending on the reader-transponder distance. For rea-
sons of compatibility, this setting can nevertheless be made. Note that the values
"0x02", "0x03" and "0x04" bring about a reduction of the power of approximately
50%.
The following settings are possible:
Value
0x02
0x03
0x04
0x05
0x06
0x07
0x08
RF600
The transmit power of the reader is set with "distance_limiting".
By default, ANT 1 is used with the preset transmit power.
0x00 = OFF
0x01 = ON
Transmit power
0.5 W
0.75 W
1.0 W
1.25 W (default)
1.5 W
1.75 W
2.0 W
Configuring the instructions
4.6 Programming the Ident profile
83