Operation

Taking Measurements

For the highest accuracy, calibrate the instrument before taking measurements
1. Create site and Data ID lists for logged data (if applicable)
2. Set the logging method (single or interval)
3. Set the Auto Stable parameters (if applicable)
4. Verify that the sensors and/or port plugs are correctly installed in all bulkhead ports
5. Install the sensor guard
6. Insert the sensors into the sample.
NOTE: Make sure to submerge the sensors completely. If using a depth sensor, submerge to where the cable
assembly attaches to the bulkhead.
7. Move the bulkhead in the sample to release any air bubbles and to provide a fresh sample to the sensors.
8. Wait for the sensor/s to stabilize in the sample.
9. On the main run screen, press
NOTE: An option to change Site and/or Data ID (if enabled) appears once
32
("Logging" on page
("Auto Stable" on page
(page
11).
ENTER
to begin logging (single or interval)
("Calibration" on page
("Logging" on page
19).
19).
26).
("Logging" on page
is pressed to begin logging.
ENTER
33).
(page
9).
19).