Operation
Figure 13 Re-Cal Prompts
Figure 14 GLP Security
18
Re-Cal Prompts
GLP
Options
Re-Cal Prompts provide a reminder to recalibrate a probe in the
user-defined number of days
The Re-Cal prompt will be displayed in the message area of the main
display when the set time has elapsed
Select the desired sensor Re-Cal prompt, then enter the desired number
of days before the Re-Cal prompt occurs.
Set the sensor value to zero (0) days (default) to turn off Re-Cal prompts.
NOTE: When enabled and the set amount of time since the last
calibration has passed, the Re-Cal prompt will be shown
when the instrument is turned on.
GLP Security
GLP
Security
The Calibration menu can be password protected to prevent accidental
or unauthorized sensor calibration (Figure 14).
1. From the GLP menu, select Security, then enter the default
password "ysi123".
2. Select Set Password [ ] and change the default password.
3. Select the Protect Cal check box to password protect the
Calibration menu.
NOTE: Write down and keep the password in a safe place. Contact
YSI Technical Support if you lose the password