Cable, Bulkhead, and Connectors

Figure 61 Cable, bulkhead,
connector maintenance

Sensor Guard

Figure 62 Sensor guard maintenance
Cable, Bulkhead, and Connector Maintenance
Wipe the bulkhead cable with a cloth dampened with a mild solution of
clean water and dish soap.
NOTICE: Install sensors or port plugs in ProDSS 4 port cable
assemblies so the bulkhead ports do not get wet when
cleaning. Exposure to water can cause damage or
corrosion to the bulkhead connectors not covered by
the warranty.
Inspect the bulkhead ports and cable connectors for contamination. If
dirty or wet, clean it with compressed air
Cable, Bulkhead, and Connector Storage
Clean the connectors and bulkhead cable. For ProDSS 4 port cables,
install the cap that protected the bulkhead during initial shipment.
Alternatively, install bulkhead port plugs when not in use
page
10).
Sensor Guard Maintenance
Remove minimal bio-fouling with a cloth soaked in a mild solution of
clean water and dish soap
Remove heavy bio-fouling by soaking the guard in a with a solution of
clean water and dish soap. Soak in vinegar to remove hard growth and
deposits.
Use a plastic scrub brush to remove any remaining bio-fouling. Rinse the
sensor guard with clean water.
NOTICE: Do not sand or polish the guard. Removal of the guard
coating can affect turbidity readings.
Maintenance and Storage
(Figure
61).
(Figure
62).
("Port plugs on