General Operation
Press "STOP" again to abort the UV run. Alternatively,
pressing "START" will cause the UV run to continue.
5.7.3
Lifetime reminder of the UV lamp
The lifetime of the UV lamp is limited to 1200 cycles. The
instrument will remind you when the UV lamp will expire by
displaying the following screen.
Beginning 50 cycles before the 1200-cycle limit, the EZ1
Advanced will display this screen every time the instrument is
switched on. "CC" indicates the number of cycles remaining.
If this number is "0", the UV lamp needs to be replaced.
Contact QIAGEN Technical Services for replacement of the
UV lamp.
5.7.4
UV lamp ignition error
If the UV lamp fails to ignite, the EZ1 Advanced will retry 2
more times. If the retries do not ignite the lamp, the following
error message appears.
Contact QIAGEN Technical Services if the lamp fails to
ignite.
5-16
Attention:
UV decontamination
not finished.
Continue:START/STOP
UV Lamp Reminder:
UV lamp expires soon
UV runs left: CC
ENT=continue
ERROR:UV Lamp
UV Lamp did not
ignite
Key:ESC
EZ1 Advanced User Manual 05/2008