4.3 Adjusting the burner
position
The burner can be positioned at three
different angles:
60°
To adjust the burner to the angle which suits
you best, simply lift it out of its base and reinsert
in the required position.
The burner is thus equally well suited for left-
handed and right-handed users.

5. Troubleshooting Guide

Fault
If the burner will
not light in spite
of the gas shutoff
valve being open
and no gas is heard
escaping.
If the burner will not
light even though
there is a loud noise
of escaping gas.
If there is a smell of
gas when the burner
is alight.
90°
120°
Cause
• Central gas supply has not
been turned on / gas bottle is
empty.
• Jets are blocked or dirty.
• Strong draught in the area of
the burner.
• Incorrect type of gas or
supply pressure.
• Connections are leaking.
• Burner is leaking.
The burner should not be used in the
90° position when working with wax,
since dripping wax may block the
jets and cause the burner to mal-
function.
Remedy
• Operational error.
• Screw off burner tubes and carefully
clean the jets with a soft cloth.
Do not screw out the jets!
• Eliminate the draught, reposition the
burner.
• Check supply pressure and type of
gas.
• Check the connections.
• If the burner leaks do not carry out
repairs yourself, send burner for
inspection.