6. Cleaning / Maintenance
The Bijou Burner requires no maintenance.
From time to time the burner should be
cleaned. Please observe the following
cleaning instructions:
Never use water or compressed air
to rinse or blow out burners! This
could flush out / blow out the sealant
grease causing the burner to leak.
1.
The maximum temperature for cleaning is
o
50
C (122ºF).
2.
Do not use cleaning agents because of the
risk of chemical reaction.
3.
To clean the valve only use a duster or
moist leather
(must not be dripping wet)!
4.
To clean the burner close the shutoff valve,
pull the hose off the burner and wipe the
burner clean with a cloth or
leather. The burner tubes can be
screwed off and cleaned separately.
5.
Do not dismantle the burner!
6.
Never drill out the burner jets.
Type of gas and power:
Rated power
Type of gas
Small
nozzle
Liquid gas
D 0.18 mm D 0.34 mm
G 30 / 50 mbar
Natural gas H
(E) G 20 / 20
D 0.32 mm D 0.55 mm
mbar
Natural gas L
(LL) G 25 / 20
D 0.35 mm D 0.61 mm
mbar
7.
8.
7. Switching to a different
The Bijou 90 conforms with the regulations set
out in DIN 30665 Part 1 and enables a switch
to another type of gas by changing the set of
burner nozzles.
Large
Load
nozzle
0.66 kW
0.66 kW
0.66 kW
Do not blow out the burner's inlet nozzle
with compressed air. This could blow away
the sealing grease and cause the burner to
leak.
The burner jets must not be screwed out
for cleaning. The burner could start to leak
and safe use is no longer
guaranteed.
If the burner nozzle becomes
plugged, disassemble the tubes than
hold the burner with the nozzle fa-
cing down. Use a hair dryer or other
device generating hot air, melt the
wax plug and allow the molten wax
to completely run out. If this method
fails to clean the nozzle, it must be
either cleaned or replaced by an
authorized technician.
type of gas
Load large
Load small
tube
tube
0.5 kW
0.16 kW
0.49 kW
0.16 kW
0.48 kW
0.16 kW
Product ref.
mark
960 F
961 N
961 N

Advertisement