Introduction PCE-AS1 is a modern air sampling meter that helps users assess microbiological pollution. The device uses the impaction principle when taking measurements. After air samples are taken and an incubation has occurred, the number of microorganisms can be determined. The air sampling meter must be used in a clean environment such as a pharmaceutical or food production facility.
Applications Pharmaceutical industry Recording of the laminar flow of microorganisms; Analysis of the cleanliness of a clean room; Testing the effectiveness of sterilization. Food industry Detection of microbial contamination in the clean room according to the HACCP concept. Hospitals Detection of pathogenic microorganisms in operating theaters, as well as in the pharmacy, storeroom or ICU.
Technical parameters • Number of micropores in the collection head: 300 • Battery: 6400 mA • Throughput: 100 l/min • Flow rate from the gathering head: 0.38 m / second (kinetic sampling) • Automatic control of the measuring volume: 10 ~ 6000 l (minimum: 10 l) •...
8 Instructions Operation diagram Main menu Start Menu Select Query of saved data Selection of memory item (max. 99 items) Function Selection via Selection via “Select” key in Selection via “Select” key “Select” key in combination with arrow keys in combination with arrow combination keys with arrow...
Welcome screen Main menu Set volume of measurement Set delay time Reset meter to factory settings Set place to save the readings Language setting Delete collected date Save settings for measurement record Backlight Save settings for measurement Date setting Time setting record * A reset to factory settings is normally not necessary.
Detailed settings To turn on the unit, press the "power" button. The indicator lights up and the device automatically performs a self-test. After the test is completed, the air sampler automatically switches to the central user interface. To specify the settings, press the "menu" button: ·...
Measuring process · Remove the cover from the collecting head. The collecting head should be sterilized in an autoclave at 121 ° C for 30 minutes (sterilization with a cloth with 75% alcohol is also possible). After sterilization, the cover should be placed back on the collecting head. The Petri dish should also be sterilized with 75% alcohol.
Maintenance and precautionary measures · Do not perform sampling with toxic corrosive gas. The device is also not for use in sour or alkaline environments. It is not allowed in areas where there is fine dust of 0.3 mg / m ³. Do not press the "sample"...