Verifying The Power Outlet Connections - Fluke Biomedical 180 User Manual

Electrical safety analyzer
Table of Contents

Advertisement

180
Users Guide
Prepare the Analyzer for use as follows:
1. To start from the same position each time, place the Analyzer's switches in
the following starting positions:
FUNCTION switch
LEAD switch
NEUTRAL switch
OUTLET switch
2. Plug the Analyzer into a properly rated outlet.
The Analyzer is equipped with a hospital-grade power plug. Grounding
reliability can be achieved only when the Analyzer is connected to an
equivalent power receptacle marked "Hospital Grade." Grounding is
important for personnel safety and to make some of the tests offered by
the Analyzer. Do not circumvent this step for any reason.

Verifying the Power Outlet Connections

Three neon lamps indicate the polarity and condition of the outlet being used,
as shown in the following chart:
REV
OK
Review all power outlets that do not indicate "correct wiring" and
have them corrected by a qualified electrician.
If the line is found faulty, correct the problem before proceeding. If the line
checks OK, plug the device to be tested into the Analyzer's line receptacle.
2-4
800.561.8187
Not applicable to isolated power systems.
Correct
Reverse
wiring
polarity
www.
V- LINE VOLTS
RL
CLOSED
OFF [center]
Note
Open
Open hot
ground
Note
.com
Hot /
Open
ground
neutral
reversed
information@itm.co